首頁 > 產品介紹 > Fully Automatic Biochemistry Analyzer, BA-A-480

Fully Automatic Biochemistry Analyzer, BA-A-480

  • Fully Automatic Biochemistry Analyzer, BA-A-480
  • Fully Automatic Biochemistry Analyzer, BA-A-480
  • Fully Automatic Biochemistry Analyzer, BA-A-480
商品圖像
  • 商品資訊

    Description

     

    DESCRIPTION

    400 Tests/Hour, Double Reagents

    Multifunctional precision sampling probe

    • External & internal mirror polish, external & internal probe washing
    • Dedicated sampling probe equipped with sensitive liquid sensor, timely feedback of reagent & sample residuals
    • Collision protection, automatic probe depth adjusting
     

    Excellent mixer design

    • Teflon coating mixer. No water dropping(reducing carry-over)
    • Excellent mixing effect with standard mixing procedure

    Reagent tray

    • 90 reagetn positions
    • Non-stop cooling system with peltier pad inside, 24 hours 2℃-14℃
    • Bar code reader(optional)

    Sample tray

    • 120 sample positions, micro cup & test tube can be used
    • Non-stop cooling system with peltier pad inside, 24 hours 2℃-14℃
    • Bar code reader(optional)

    Reaction tray

    • 90 reaction cuvettes
    • Reaction volume as low as 150μl
    • Stable & accurate temperature (37±0.1℃)for reaction

    Washing systerm

    • 2*6 steps automatic washing
    • Efficient cleaning to ensure accurate results and reliable diagnnosis
     

    Advanced operating system

    • User friendly software interface simple, easy-to-operation
    • Sample automatic dilution(decrease, increase, normal)
    • Reaction result automatic checking and warning, calibaration result automatic checking, warning alarm log
    • Linear & nonlinear calibration, Multi-points calibration up to 8 points
    • Quality control apply to westgard, L-T plot, cumulative
     

    SPECIFICATIONS

     
    General information
    Machine type Random access, open reagent & locked reagent (optional)
    Test speed Constant 400 tests/hour (Mono reagent or double reagent)
    Test principle Colorimetric method, turbidimetry
    Test method 1 point end, 2 point end, Fixed-Time, Kinetics
    Calibration type Linear & nonlinear
    Sample unit
    Sample tray 120 sample positions
    Sample tube Micro cup & test tube & blood collection tube can be used.(φ12 -13 )mm * (25 -100 )mm
    Sample volume 2-30µL step by 0.1µL
    Sample probe Liquid surface detection, timely tracking for liquid volume.Collision avoidance
    Sample barcode Sample barcode reader (optional)
    Reagent unit
    Reagent tray 90 reagent positions, Non-stop cooling system with peltier pad inside, 24 hours 2°C-14°C, open system, could accept any reagent
    Reagent volume 20-300µL, step by 1µL
    Reagent needle Liquid surface detection, timely tracking for liquid volume.Collision avoidance.
    Regent barcode Regent barcode reader (optional)
    Reaction unit
    Reaction tray 90 reaction cuvettes, new amorphous optical plastics reaction cups, perfect transmittance
    Reaction volume 150µL-330µL
    Reaction time 10 minutes
    Temperature Peltier pad incubation system, reaction temperature (37°C ±0.1 °C)
    Mixing system
    Mixer paddle-type mixing, effective reducing of carry-over
    Optical system
    Light source Halogen lamp
    Wavelength 340nm; 405nm; 450nm; 505nm; 546nm; 578nm; 630nm; 700nm(4 more options)
    Absorbance range 0-4.0 Abs
    Resolution 0.0001 Abs
    Operating system
     Operation system Windows XP, Windows 7, Windows 8, Windows 10
    Data processing Automatic calibration, test combination, effective time management of reagents, serum index, whole process detection, linear expansion of enzyme, blank deduction,dirty cuvette memory , cross infection prevention, patient information memory and association input, automatic report audit, data fuzzy query, report statistics and printing, reference range grading, alarm information classification,operation privilege classification management
    Report printing 6 formats optional, support custom pattern
    Interface TCP/IP network interface
    Working conditions
    Power supply AC 110/220V ± 10%, 50/60 Hz 650W
    Water consumption 15L/hour(maximum)
    Dimension 1110mm * 765mm * 1150mm(L*W*H)
    Weight 200kg
    Language All languages
     
     